How do you choose battery backup for a CPAP machine?

Start with the exact CPAP model and the equipment your sleep clinic supplied. A battery that can power one machine for the night may fall short with a heated tube and humidifier. Ask your sleep service or equipment supplier to confirm a compatible battery and what to do in an outage.

What to check before buying

  1. Read the model name on the machine. For example, ResMed lists separate DC converters for the AirSense 10 and AirSense 11. Ask which power supply or converter is approved for your particular device.
  2. Tell the clinic or supplier which accessories you use and how long the backup has to last. Ask if their suggested option changes over automatically or needs to be connected by hand.
  3. Check whether you can recharge between nights. If you depend on electrically powered medical equipment, ask your electricity network operator about its Priority Services Register and what help is available.

ResMed UK power accessories · University Hospitals of North Midlands: CPAP power-cut advice · Ofgem Priority Services Register guidance

How much difference does a heated tube or humidifier make?

In ResMed’s AirSense 10 DC converter table, these three examples use the same test pressure of 10 cm H₂O. The table shows how the accessories change the energy needed over eight hours. The test settings are examples. Keep the treatment settings prescribed for you and ask your clinic about any changes.

Calculated energy at the 12 V source from ResMed’s published current figures
AirSense 10 setup in ResMed’s tableAverage currentEnergy over 8 hours
SlimLine tube0.93 A89 Wh
SlimLine tube and humidifier at 42.30 A221 Wh
ClimateLineAir heated tube and humidifier on Auto3.77 A362 Wh

We multiplied 12 V by the published current and by eight hours. These are estimates of energy drawn at the source with the matching converter. ResMed’s 2018 guide states a 50% safety margin for its eight-hour deep-cycle battery sizing tables. For these rows it lists 12, 28 and 46 Ah respectively at 12 V, equivalent to nominal battery energies of 144, 336 and 552 Wh. Those are the guide’s battery sizes, not portable power-station recommendations. Ask your supplier to confirm a current compatible system, including its losses and reserve. The figures apply to the AirSense 10 example and its DC setup. For other models or mains power, use the relevant manufacturer guidance. ResMed battery guide, Air10 converter table (PDF)

If the power goes off during treatment

University Hospitals of North Midlands advises its CPAP patients who become aware of a power cut to remove the mask and switch the machine off. Follow the outage plan from your own sleep service and contact it if treatment cannot resume. Do not assume advice for obstructive sleep apnoea applies to a ventilator or another form of respiratory support. Read the trust’s advice

Check automatic shut-off before relying on a battery

A power station may turn off a lightly loaded output after a timeout. Ask the supplier to confirm the appropriate settings and test the complete approved setup, including humidifier and heated tube, before relying on it overnight. Disabling an energy-saving timeout can increase the station’s own energy use; include that in the budget.

What to ask about a portable power station

If your supplier confirms a portable station is suitable, check the actual runtime with your power supply and accessories, the station’s low-load shut-off behaviour and what happens when mains power fails or returns. A station’s capacity label gives you a starting point. Confirm the complete setup and your outage plan with your care team before relying on it.

Correction, 16 November 2026: Named the NHS trust, added its outage advice and explained the source’s 50% sizing margin and the limits of that older guidance.

Can I size the battery from the adaptor rating?

The adaptor’s maximum output is not a measured overnight energy total. Use model-specific guidance from the equipment supplier, including the humidifier and heated tube settings you require. Do not change prescribed therapy to fit a smaller battery.

Will a large battery always stay on overnight?

No. Automatic low-load shut-off can interrupt a small load even with energy remaining. Confirm the relevant settings and test the approved machine, adaptor and battery arrangement before relying on it.

Updated